FDA rules

21 C.F.R. § 890.5870

Non-invasive bone growth stimulator.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. A non-invasive bone growth stimulator is a prescription device that provides stimulation through electrical, magnetic, or ultrasonic fields. The device is intended to be used externally to promote osteogenesis as an adjunct to primary treatments for fracture fixation and spinal fusion or as a treatment for established nonunions or failed fusions.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Clinical data must demonstrate that the device performs as intended under anticipated conditions of use. Imaging data must demonstrate fusion at the treatment site.

(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Critical performance and safety characteristics of the device, considering the operational modality of the device, must be verified and validated to ensure:

(i) Intended design outputs are delivered to the patient;

(ii) Thermal safety and thermal reliability;

(iii) Signal characteristics are within safe physiologic limits; and

(iv) Device reliability is consistent with the expected use-life.

(3) Patient-contacting components of the device must be demonstrated to be biocompatible.

(4) Performance data must demonstrate the electrical safety and electromagnetic compatibility of the device.

(5) Appropriate software verification, validation, and hazard analysis must be performed.

(6) Labeling comprehension testing must demonstrate the patient can correctly use the device based solely on reading the instructions for use.

(7) Labeling for the device must include the following:

(i) Warning against use on compromised skin or when there are known skin sensitivities;

(ii) Appropriate warnings for patients with implanted medical devices;

(iii) A detailed summary of the supporting clinical data, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;

(iv) A clear description of the device;

(v) Instructions on appropriate usage, duration, and frequency of use;

(vi) Instructions for maintenance and safe disposal;

(vii) Validated instructions for appropriate cleaning of any reusable components;

(viii) Specific warnings regarding user burns, electrical shock, and skin irritation; and

(ix) The risks and benefits associated with use of the device when used as intended.

Source and amendment notes

Amendments

[91 FR 20362, Apr. 16, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder