21 C.F.R. § 888.5895

Wearable vibration device for orthopedic use.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. A wearable vibration device for orthopedic use is a wearable device that uses mechanical vibrations, targeted to specific regions of the skeleton, to reduce loss of bone strength or bone mineral density.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must evaluate the reduction in loss of bone strength or bone mineral density. Testing must evaluate patient risks, including regional pain, dizziness, blurred vision, muscle weakness, difficulty with balance, headache, and nausea. The testing must include a minimum of 1 year of follow-up.

(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be provided:

(i) Characterization of the designed outputs of the device;

(ii) Demonstration of the reliability and reproducibility of device output; and

(iii) Validation that vibration characteristics received by the patient are within safe physiologic limits.

(3) The patient-contacting components of the device must be demonstrated to be biocompatible.

(4) Performance data must be provided to demonstrate the electrical safety, mechanical safety, thermal safety, battery safety, and electromagnetic compatibility of the device.

(5) Software verification, validation, and hazard analysis must be performed.

(6) Human factors or usability evaluation must demonstrate that the user can correctly use the device, based solely on reading the directions for use.

(7) Labeling must include the following:

(i) A summary of the clinical performance testing with the device, including clinical outcomes and observed adverse events;

(ii) Information regarding limitations of the clinical significance of the clinical performance testing;

(iii) Instructions on duration and frequency of use, and when to discontinue use of the device; and

(iv) Instructions for device maintenance, including appropriate cleaning of any reusable components and safe disposal of the device.

Source and amendment notes

Amendments

[91 FR 69694, Nov. 5, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder