21 C.F.R. § 878.4405

Focused ultrasound system for non-thermal, mechanical tissue ablation.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. This device uses focused ultrasound to mechanically ablate soft tissue. The device is not intended to thermally ablate tissue.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must document the adverse event profile and characterize tissue destruction.

(2) Animal performance testing must demonstrate that the device mechanically destroys targeted tissue while characterizing the chronic safety profile, including thermal and mechanical injury to adjacent, non-target tissue.

(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

(i) Acoustic characterization at clinically relevant settings;

(ii) Determination of the minimum drive voltages necessary to sustain a bubble cloud in the target tissue;

(iii) Validation of mechanisms to prevent energy delivery that would result in adverse thermal effects;

(iv) Availability of real-time monitoring during the procedure;

(v) Validation that the treatment zone is limited to the defined target tissue; and

(vi) Validation of mechanisms to prevent, pause, and terminate ablation in the event of device failure.

(4) Performance data must support the electrical safety and electromagnetic compatibility of the device.

(5) All patient-contacting components of the device must be demonstrated to be biocompatible.

(6) Software validation, verification, and hazard analysis must be performed.

(7) Human factors testing must demonstrate that the user can safely and correctly use the device.

(8) Labeling must include the following:

(i) A warning that focused ultrasound ablation should only be considered in patients with sufficient functional reserve to withstand the destruction of the planned volume of tissue;

(ii) A statement that the device has not been evaluated for the treatment of any specific disease or condition; and

(iii) A detailed summary of the clinical testing with the device.

Source and amendment notes

Amendments

[91 FR 63495, Oct. 6, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder