(a) Identification. An implanted tibial electrical urinary continence device is an implanted prescription device that receives power from a non-implanted external power source to provide electrical stimulation of the tibial nerve in proximity to the ankle. The device is intended for the treatment of overactive bladder related symptoms of urge urinary incontinence, urinary urgency, urinary frequency, and nocturia.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following testing must be conducted:

(i) Electrical performance testing of the device must be conducted to validate the specified electrical output and duration of stimulation of the device; and

(ii) Testing must verify the implant can withstand clinically relevant forces during and after implantation.

(2) The patient-contacting components of the device must be demonstrated to be biocompatible.

(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.

(4) Performance data must support the shelf life of the device by demonstrating continued sterility of patient-contacting components, package integrity, and device functionality over the identified shelf life.

(5) Performance testing must demonstrate the electromagnetic compatibility, electrical safety, thermal safety, and wireless performance of the device.

(6) Software verification, validation, and hazard analysis must be performed.

(7) Performance testing must evaluate the compatibility of the device in a magnetic resonance environment.

(8) Labeling for the device must include:

(i) A contraindication against use during pregnancy;

(ii) A contraindication against using the device in men who have Benign Prostatic Hyperplasia (BPH) or other lower urinary tract obstructions;

(iii) A detailed summary of the device technical parameters and the typical course of treatment;

(iv) Device- and procedure-related adverse events pertinent to use of the device; and

(v) A shelf life for any sterile components.

(9) Patient labeling must include:

(i) Post-operative care instructions to avoid infection and inflammation of the surgical site;

(ii) Instructions to avoid overstimulation related nerve and tissue damage;

(iii) Instructions to avoid mechanical injuries to nerve and tissue caused by the implanted component;

(iv) Instructions for reprocessing/cleaning of any reusable components;

(v) Clinical performance reported by relevant subgroups;

(vi) The risks and benefits associated with use of the device;

(vii) Information on the typical course of treatment;

(viii) Instructions to avoid pain and discomfort; and

(ix) Instructions to avoid interference with other medical devices.

Source and amendment notes

Amendments

[91 FR 57506, Sept. 10, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder