FDA rules

21 C.F.R. § 870.2720

Hyperoxia monitoring device adjunct to pulse oximetry.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. A hyperoxia monitoring device adjunct to pulse oximetry is a device that monitors elevated hemoglobin oxygen saturation levels as an adjunct to arterial oxygen saturation monitoring.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Clinical performance testing under anticipated conditions of use must evaluate the accuracy of hyperoxia monitoring by the device and evaluate all adverse events.

(2) Human factors/usability testing must demonstrate that the user can correctly use the device, based solely on reading the instructions for use.

(3) Performance data must demonstrate the electromagnetic compatibility and electrical safety of the device.

(4) The patient-contacting components of the device must be demonstrated to be biocompatible.

(5) Performance testing must validate the reprocessing instructions for the device, including demonstration of device performance after repeated cleaning and disinfection.

(6) Software verification, validation, and hazard analysis must be performed. Performance testing must demonstrate compatibility with pulse oximeter devices labeled to be compatible with the device.

(7) Labeling must include:

(i) Cleaning and disinfection instructions;

(ii) A summary of the clinical performance testing with the device;

(iii) A warning against over-reliance on device output without using arterial blood gas values for confirmation; and

(iv) Instructions to monitor oxygen delivery and patient clinical/cardiovascular status when device output changes.

Source and amendment notes

Amendments

[91 FR 63493, Oct. 6, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder