21 C.F.R. § 213.165

Testing and release for distribution.

Current through eCFR issue date November 12, 2026.

Sections

(a) For each batch of medical gas, there shall be appropriate laboratory determination of satisfactory conformance to final specifications for the medical gas, including the identity and strength, prior to release.

(b) Any sampling and testing plans shall be described in written procedures that shall include the method of sampling, the number of units per batch to be tested, and acceptance criteria. Such written procedures shall be followed.

(c) The accuracy, sensitivity, specificity, and reproducibility of test methods employed by the firm shall be established and documented. Such validation and documentation may be accomplished in accordance with § 213.194(a)(2). The suitability of all testing methods shall be verified under actual conditions of use.

(d) Medical gases failing to meet established standards or specifications and any other relevant quality criteria shall be rejected.

(e) This section does not apply to the filling of a designated medical gas or medically appropriate combination of designated medical gases via liquid to liquid into a container at a delivery site.

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder